MC-B70
The coil is suitable for focused stimulation. It is slightly bend to closely follow curved shapes. Equipped with trigger button to support easy operation. Usable with a MagPro Compact stimulator with converter.
Product details
Product sheetPerformance
Number of stimulations before warm-up at ambient temperature 20°C:| Mean output 75% of maximum at 1pps. | 400 pulses |
Mechanical Properties
Weight of transducer head | 1.1 kg /2.4 lbs |
Cable length | 1.7 m /5.58 ft. |
Dimensions of transducer head | 169 x 112 x 16/33 mm /6.65 x 4.41 x 0.63/1.3in |
Angle | 150° |
Coil winding data
Inner diameter | 27 mm /1.06 in. |
Outer diameter | 97 mm /3.7 in. |
Winding height | 6 mm/0.24 in. |
Number of windings | 2 x 10 |
Magnetic and electrical properties
Max initial dB/dt | 31 kT/s near the coil surface |
Active pulse width | 280µs (Biphasic) |
Operating period
5 years
Ordering number
9016E056-