 Number of stimulations, before warm-up at ambient temperature 20°C (68°F):| Mean output 75% of maximum at 1pps.  5,500|pulses
 Protocol: 60 trains @ 50 pulses/train @ 10pps @ Inter Train Interval: 25s @ Output=75% @ total number of stimulations 3,000  

One protocol can be performed without overheating the coil.

Minimum cooling time between protocols: 2.5 hours at 20°C/68°F or 1 hour at 7°C/45°F

Mechanical Properties

Weight of transducer head  2.5 kg/5.5 lbs 
 Cable length  2 m/6.6 ft
 Dimensions of transducer head  180x116x 50/67 mm /7.1 x 4.6 x 1.8/2.5 in
 Angle  150°

Coil winding data

Inner diameter  23 mm /0.9 in 
 Outer diameter  97 mm / 3.8 in
 Winding height  12mm / 0.5 in
 Number of windings  2 x 11

Magnetic and electrical properties

Max initial dB/dt (near the coil surface)  28 kT/s

 

Active pulse width (Biphasic)

280µs

Operating period

5 years

Ordering number

9016E040-

Users in the USA

Caution: investigational device. Limited by Federal law to investigational use. In accordance with US Federal regulations, and IDE and/or IRB approval may be required.