World Leading Magnetic
Stimulation Systems

MagPro is a complete line of non-invasive Magnetic Stimulation Systems, including both dedicated and general-purpose stimulators.

 
 

MRi-B88

Welcome to MagVenture - Magnetic Adventures

MRi-B88 TMS Solution for fMRI - for use with MRI scanners up to 4 Tesla
For more information download this PDF file or contact MagVenture



The TMS Solution for MRI consists of the following items:

  • Coil MRi-B88
  • Emergency Stop
  • Remote Control
  • Power Line Filter
  • Cushion

 Coil MRi-B88
MRi-B88

 Coil with Cable
Cable for MRi-B88

 Emergency Stop
Emergency Stop

Power Line Filter 
Power Line Filter

 Remote Control
Remote Control

 Cushion
Cushion

 

 

 

· The coil is suitable for focused stimulation

· Designed for use in MRI scanners up to 4 Tesla with biphasic Waveform Standard mode. Not for use with power mode or any other waveforms.



Mechanical Properties

Weight of transducer head

1.8 kg

Dimensions of transducer head

200 x 125 x 50 mm

Cable length *)

6.2m

Coil Winding Data

Inner diameter

2 x ø43

Outer diameter

2 x ø90

Winding height

15 mm

Number of windings

2x9

Magnetic and Electrical Properties

Max initial dB/dt

≥34 kT/s

Max B Field 1.8 Tesla

Active pulse width

280µs (Biphasic)

Performance

Number of stimulations before warm-up at ambient temperature 20°C:
Mean output 75% of maximum at 1pps.

1000 pulses

Ordering Number

9016E0661 Coil MRi-B88
9016C0741 Remote Control and Emergency stop
9016C0751 Power Line Filter
9016C0761 Extra Cushions (set of 5)
 

*) The length of all cables used in the TMS Solution for MRI is customer specific


 