| Technical Data |
|
Dimensions |
(HxWxD:) 184 x 94 x 40mm |
Weight |
0.7kg |
Number of Inputs |
1 input protected against electro-static discharge. Balanced inputs. 1 pc. 5-pole DIN 240° connector for electrode cable |
Sound |
Output for external loudspeaker, 3.5mm jack. |
Patient Safety |
EMG channel galvanically isolated 1.5 kV RMS |
Input Impedance |
200 MΩ // 100 pF (balanced), >1000 MΩ // 50 pF (common mode) |
Noise Level |
Typical 0.6 µVrms at bandwidth 2 Hz to 20 kHz and shorted input |
Common Mode Rejection Ratio |
From surface electrode, through cable and amplifier: >55 dB. Direct: >100 dB |
Isolation Mode Rejection Ratio |
From input to power ground: >160 dB |
Sensitivity Factors |
100, 200, 500 µV/Div, 1, 2, 5, 10 mV/Div |
Time Scales |
1, 2, 5, 8, 10 ms/Div |
Trigger Mode |
Level, Autotrig on stim |
Lower Frequency Limits (-3dB): DSP |
1, 2, 5, 10, 20, 50, 100 Hz, |
Upper Frequency Limits (-3dB): DSP |
1, 2, 5, 10, 20 kHz, |
Anti Aliasing |
20 kHz (-3 dB), 1st order |
Sampling |
100 ks/s, 16 bit |
| |
|
| Ordering Number |
9016C070- | Download datasheet (PDF) Download Accessories Catalogue (PDF) |
|

- 1 channel EMG amplifier to be mounted on the back of the MagPro system.
- Measurement of Motor Evoked Potentials (MEP).
- Specially designed for determination of Motor Threshold.

|
|